Soft-db Opus Suite Building Acoustics Module Airborne Soun Manual de usuario Pagina 16

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 40
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 15
15
6.1 Sensor Calibration
The input sensitivity can be calibrated using the calibration function and a sensor calibrator. Click the
button on the Setup interface to launch the Calibration interface.
Step 1 Adjust the calibration parameters
The defaults values are:
Averaging time: 5 s
Frequency: 1 kHz
Calibrator Level: 94 dB
Step 2 Install the calibrator device on the microphone
Step 3 Click START
After the average time is elapsed, the sensitivity value will update.
Step 4 Click OK to accept the sensitivity value
Vista de pagina 15
1 2 ... 11 12 13 14 15 16 17 18 19 20 21 ... 39 40

Comentarios a estos manuales

Sin comentarios